Section 8.3 — Nucleic Acid Detection and Quantitation in Solution

Printer-Friendly

Nucleic Acid Detection and Quantitation in Solution

Through intensive research efforts in both chemical synthesis and bioassay development, Molecular Probes scientists have developed rapid and exceptionally sensitive fluorescence-based assays for quantitation of nucleic acids in solution. The Quant-iT Assay Kits represent the most advanced quantitation systems for DNA, RNA or protein samples. Described in detail below, these state-of-the-art assays deliver high sensitivity and specificity together with a streamlined protocol, prediluted standards and a ready-to-use buffer.

In addition, the PicoGreen dsDNA quantitation reagent, OliGreen ssDNA quantitation reagent and RiboGreen RNA quantitation reagent — optimized for double-stranded DNA, oligonucleotides and RNA, respectively — have a high affinity for nucleic acids and an extremely large fluorescence enhancement upon binding, making possible the direct detection of minute amounts of nucleic acids in complex solutions within minutes, usually without interference from other biomolecules. These Patented reagents and quantitation assays provide the following advantages:

  • Sensitivity. The PicoGreen dye–, OliGreen dye– and RiboGreen dye–based fluorescence assays are up to 10,000-fold more sensitive than UV absorbance measurements and at least 400-fold more sensitive than assays that use the Hoechst 33258 dye ref (H1398, H3569; FluoroPure Grade, H21491), requiring much less sample for quantitation.
  • Accuracy. Unlike measurements of UV absorbance, these assays are not affected by the presence of proteins, free nucleotides or very short oligonucleotides, making quantitation of intact oligonucleotides and nucleic acids much more accurate in complex mixtures such as serum or whole blood.
  • Precision. The average standard deviations of triplicate assays using these reagents are typically less than 5%.
  • Simplicity. These assays have a very simple protocol that requires no separation steps, making them ideal for automated, high-throughput measurements.
  • Broad dynamic range. Quantitation is accurate over four orders of magnitude for the PicoGreen and OliGreen assays, with a single dye concentration. The RiboGreen assay is accurate over three orders of magnitude.
  • Instrument compatibility. Quantitation assays can be performed using a fluorescence microplate reader with standard filters optimized for fluorescein-like dyes, a relatively inexpensive filter-based spectrofluorometer or a standard spectrofluorometer.
  • Convenience. Each reagent is available separately or in a kit that additionally contains a nuclease-free buffer and standards. The PicoGreen dsDNA and the RiboGreen RNA quantitation assays are also available in the RediPlate 96 format — a prepared 96-well microplate with standards included — for high-throughput applications.

Quant-iT Assay Kits for DNA and RNA

The Quant-iT family of assay kits provides state-of-the-art reagents for sensitive and selective quantitation of DNA, RNA or protein samples using a standard fluorescence microplate reader (Table 8.12). These kits have been specially formulated with ready-to-use buffers, prediluted standards and easy-to-follow instructions, making quantitation both accurate and extremely easy (Figure 8.52). Each Quant-iT assay is:

  • Ready to use. Only the dye is diluted in the supplied buffer; dilution of standards or buffer is not required.
  • Easy to perform. Just add the sample to the diluted dye and read the fluorescence.
  • Highly sensitive. The Quant-iT protein assay is orders of magnitude more sensitive than UV absorbance measurements.
  • Highly selective. Separate kits are available for quantitating DNA, RNA (see below) or protein (Section 9.2), with minimal interference from common contaminants.
  • Precise. CVs are generally less than 5%.

Because the fluorescent dye in each Quant-iT Kit matches common fluorescence excitation and emission filter sets in microplate readers, these assay kits are ideal for high-throughput environments, as well as for small numbers of samples.

Quant-iT DNA Assay Kits

The Quant-iT DNA Assay Kits simplify DNA quantitation without sacrificing sensitivity. The Quant-iT DNA High-Sensitivity DNA Assay Kit (Q33120) provides a linear detection range between 0.2 ng and 100 ng double-stranded DNA (dsDNA) (Figure 8.53), corresponding to initial experimental sample concentrations between 10 pg/µL and 100 ng/µL. This high-sensitivity DNA assay is ideal for quantitating PCR products, viral DNA, DNA fragments for subcloning and other applications requiring small amounts of DNA. The Quant-iT DNA Broad-Range DNA Assay Kit (Q33130) provides a linear detection range between 2 ng and 1000 ng double-stranded DNA (dsDNA) (Figure 8.54), corresponding to initial experimental sample concentrations between 100 pg/µL and 1000 ng/µL. This broad-range DNA assay eliminates the need to dilute concentrated samples, such as genomic DNA and miniprep DNA, prior to high-throughput procedures. Both Quant-iT DNA assays are highly selective for dsDNA over RNA, and have proven to be accurate even in the presence of an equal mass of RNA. Moreover, their fluorescence signals are unaffected by many common contaminants, including free nucleotides, salts, solvents and proteins. Each Quant-iT DNA Assay Kit contains:

Sufficient reagents are provided to perform 1000 assays, based on a 200 µL assay volume in a 96-well microplate format; this assay can also be adapted for use in cuvettes or 384-well microplates. Both the high-sensitivity assay and the broad-range assay can be detected using standard fluorescein filters, and the fluorescence signal is stable for three hours at room temperature. The Quant-iT HS reagent is a new formulation of Molecular Probes' PicoGreen reagent, which is described below.

Quant-iT RNA Assay Kit

The Quant-iT RNA Assay Kit (Q33140) provides the first homogeneous assay ever developed for quantitating RNA in the presence of DNA. This RNA assay exhibits a linear detection range between 5 ng and 100 ng RNA (Figure 8.55), corresponding to initial experimental sample concentrations between 250 pg/µL and 100 ng/µL. Because of the high selectivity of the Quant-iT RNA reagent for RNA over dsDNA, this assay provides accurate RNA quantitation even in the presence of an equal mass of DNA. The fluorescence signal is unaffected by many common contaminants, including free nucleotides, salts, solvents and proteins, making this assay ideal for measuring samples for microarray, RT-PCR and Northern blot procedures. Each Quant-iT RNA Assay Kit contains:

  • Quant-iT RNA reagent
  • Quant-iT RNA buffer
  • A set of eight prediluted Escherichia coli rRNA standards between 0 and 10 ng/µL
  • Easy-to-follow instructions (Quant-iT RNA Assay Kit)

Sufficient reagents are provided to perform 1000 assays, based on a 200 µL assay volume in a 96-well microplate format; this assay can also be adapted for use in cuvettes or 384-well microplates. The fluorescence signal exhibits excitation/emission maxima of 644/673 nm and is stable for three hours at room temperature.

PicoGreen dsDNA Quantitation Assay

PicoGreen dsDNA Quantitation Reagent and Kits

The Patented PicoGreen reagent ref (P7581, P11495) and Kits (P7589, P11496) can accurately quantitate as little as 25 pg/mL of dsDNA in a fluorometer or 250 pg/mL (typically 50 pg in a 200 µL volume) in a fluorescence microplate reader. The PicoGreen assay is more than 10,000 times as sensitive as conventional UV absorbance measurements at 260 nm (an A260 of 0.1 corresponds to an ~5 µg/mL dsDNA solution) and at least 400 times more sensitive than the Hoechst 33258 dye–based assay. It is even more sensitive than assays based on our YO-PRO-1 and YOYO-1 dyes, which have reported detection limits of approximately 2.5 ng/mL ref and 0.5 ng/mL,ref respectively. Although the PicoGreen reagent is not specific for dsDNA, it shows a >1000-fold fluorescence enhancement upon binding to dsDNA, and much less fluorescence enhancement upon binding to single-stranded DNA (ssDNA) or RNA, making it possible to quantitate dsDNA in the presence of equimolar amounts of ssDNA, RNA or proteins ref (Figure 8.56). The PicoGreen reagent also selectively detects DNA–RNA hybrids in the presence of ssDNA and RNA.ref Differences in the emission lifetimes of the PicoGreen complexes with dsDNA and ssDNA make it possible to quantitate the relative amounts of each species in solution by time-resolved measurements.ref By contrast, UV absorbance measurements cannot distinguish between dsDNA, ssDNA and RNA or proteins. Thus, the PicoGreen reagent allows direct quantitation of PCR amplicons without purification from the reaction mixture and makes it possible to detect low levels of DNA contamination in recombinant protein products.ref In comparison to the Hoechst 33258 dye, which shows significant AT selectivity, the PicoGreen reagent shows little if any AT- or GC-selectivity and is thus accurate for quantitating DNA from almost any source. The PicoGreen reagent can be excited at 488 nm with an argon-ion laser, and is reported to be the best nucleic acid stain for analysis of single DNA molecules in a flow cytometer.ref A method that utilizes the PicoGreen reagent for the absolute quantitation of cDNA using real-time PCR has been reported.ref

The protocol for the PicoGreen assay is easy to conduct (Quant-iT PicoGreen(R) dsDNA Reagent and Kits) and requires very few steps — the dye is simply added to the sample and incubated for five minutes, then the fluorescence is measured (Figure 8.57). In addition, the fluorescence signal from binding of the PicoGreen reagent to dsDNA is linear over at least four orders of magnitude (Figure 8.58) with a single dye concentration, whereas assays using ethidium bromide, Hoechst 33258 or the YOYO-1 dye exhibit a much more limited linear range.ref We have found that this linearity is maintained in the presence of several compounds commonly found in nucleic acid preparations, including salts, urea, ethanol, chloroform, detergents, proteins and agarose ref (Table 8.13). The PicoGreen assay is faster, more precise and much less expensive than dipstick or blot assays.

The PicoGreen assay is useful for quantitating DNA templates for PCR,ref labeling reactions, electrophoretic mobility-shift (bandshift) assays, DNA-footprinting assays and filter-binding assays, and for measuring yields from PCR reactions,ref DNA minipreps and maxipreps, cDNA synthesis and nuclease protection assays. The simplicity and selectivity of the assay also make it ideal for high-throughput automated quantitation assays used in forensic and genomics research. Furthermore, the PicoGreen reagent has been used for:

  • Genotyping by allele-specific PCR ref
  • Quantitating dsDNA samples before and after PCR amplification or after agarose gel electrophoresis ref
  • Determining PCR amplification yields before sequencing ref
  • Automating quantitation of DNA isolated from biological samples or obtained from PCR reactions, prior to running DNA typing gels for high-throughput genotyping ref
  • Quantitating DNA from buccal scrapes prior to DNA profiling by short tandem repeat (STR) analysis ref
  • Identifying contaminating DNA in recombinant protein products ref or purified monoclonal antibody preparations ref
  • Monitoring DNA strand breaks in plasmids and bacterial artificial chromosomes ref
  • Quantitating the efficiency of DNA extraction from frozen and formalin-fixed tissue sections ref
  • Detecting mammalian telomerase activity in tumor cells using the PCR-based TRAP assay ref
  • Developing assays for DNA polymerase ref
  • Measuring the activity of reverse transcriptase ref or DNase I ref
  • Measuring supercoiled DNA forms in solution, based on their renaturation properties ref
  • Monitoring plasmid production during fermentation and downstream processing ref
  • Quantitating DNA denaturation as a measure for DNA damage in purified DNA preparations, cell lysates or homogenized solid tissues ref

Each vial of the PicoGreen dsDNA quantitation reagent (P7581) contains a sufficient amount of dye for at least 200 assays using a 2 mL assay volume and a standard fluorometer, or 2000 assays using a 200 µL assay volume and a fluorescence microplate reader. The product is accompanied by a simple protocol that ensures linear and reproducible quantitation of dsDNA. We also provide the PicoGreen reagent in a kit (P7589) that contains:

  • PicoGreen dsDNA quantitation reagent
  • Low-fluorescence, nucleic acid– and nuclease-free assay buffer concentrate, essential for the accurate measurement of dsDNA
  • dsDNA standard solution for assay calibration
  • A detailed protocol for dsDNA quantitation (Quant-iT PicoGreen(R) dsDNA Reagent and Kits)

This kit provides sufficient reagents for 200 assays using a 2 mL assay volume and a standard fluorometer or 2000 assays using a 200 µL assay volume and a fluorescence microplate reader. Both the stand-alone reagent and the kit are also available in special packaging, in which the PicoGreen reagent is supplied as 10 vials of 100 µL aliquots for added convenience (P11495; Kit, P11496). The special packaging reduces thawing times, provides individual aliquots for each person performing the assay, and allows smaller amounts of dye to be taken into the field for analysis of water or other samples.

RediPlate 96 PicoGreen dsDNA Quantitation Kits and Microplate

The PicoGreen assay is also available as the convenient RediPlate 96 PicoGreen dsDNA Quantitation Kits and microplate (R21495, R32716, R32701), in which the reagent is predispensed into a 96-well microplate. Nuclease-free buffer and the sample are simply added to the microplate wells — there is no need to handle the PicoGreen reagent. After a five-minute incubation, the microplate is ready to read in a fluorescence microplate reader. The RediPlate PicoGreen assay has sensitivity and linearity identical to that of the PicoGreen reagents and kits described above (Figure 8.59).

We offer two different RediPlate 96 PicoGreen dsDNA Quantitation Kits, as well as a RediPlate 96 PicoGreen dsDNA quantitation microplate:

  • In the RediPlate 96 PicoGreen dsDNA Quantitation Kit (R21495), the microplate is provided in a resealable foil packet, and it snaps apart into twelve strips to permit assays in any multiple of eight (Figure 8.60). Eleven of the strips are preloaded with the PicoGreen reagent; the remaining strip, marked with blackened tabs, contains a series of DNA standards for generating a standard curve. This kit also includes nuclease-free buffer and detailed instructions (RediPlate 96 PicoGreen(R) dsDNA Quantitation Kit).
  • In the RediPlate 96 PicoGreen dsDNA Quantitation Kit (R32716), the microplate is provided in a resealable foil packet and snaps apart into twelve strips, but in this microplate all twelve strips are preloaded with the PicoGreen reagent; the DNA standards are not predispensed into any of the microplate wells. Instead, a λ DNA standard is provided in a separate vial for generating a standard curve in the appropriate experimental range. Nuclease-free buffer and detailed instructions (RediPlate 96 PicoGreen(R) dsDNA Quantitation Kit) are also provided.
  • The RediPlate 96 PicoGreen dsDNA quantitation microplate (R32701, Product Information Sheet) is the same microplate provided in the RediPlate 96 PicoGreen dsDNA Quantitation Kit (R32716) with all wells preloaded with PicoGreen reagent, for those who wish to prepare their own nuclease-free buffer and DNA standard curve. Multiple plates are available at significant discounts.

OliGreen ssDNA Quantitation Reagent and Kit

For researchers working with ssDNA and oligonucleotides, as well as for companies that synthesize oligonucleotides, Molecular Probes offers our Patented OliGreen ssDNA quantitation reagent (O7582) and Kit (O11492). Short, synthetic oligonucleotides are used in a number of molecular biology techniques, including DNA sequencing, site-directed mutagenesis, DNA amplification, antisense gene suppression and in situ and blot hybridization. The conventional methods for quantitating oligonucleotides are not very sensitive, often requiring highly concentrated samples, and are quite subject to interference from sample contaminants. The most commonly used technique for measuring oligonucleotide and ssDNA concentrations is the determination of absorbance at 260 nm (an A260 of 0.1 corresponds to an ~3 µg/mL solution of a synthetic 24-mer M13 sequencing primer).

The OliGreen ssDNA quantitation reagent enables researchers to routinely quantitate as little as 100 pg/mL of ssDNA or oligonucleotide (200 pg in a 2 mL assay volume with a standard fluorometer) or 200 pg in a 200 µL assay volume using a fluorescence microplate reader (Figure 8.61). Thus, quantitation with the OliGreen reagent is about 10,000 times more sensitive than quantitation with UV absorbance methods and at least 500 times more sensitive (and far faster, with a greater throughput) than detecting oligonucleotides on electrophoretic gels stained with ethidium bromide. Using an easy-to-follow protocol (Figure 8.57, Quant-iT OliGreen(R) ssDNA Reagent and Kit) and fluorescein excitation and emission wavelengths, we have quantitated oligonucleotides that range from 10 to 50 nucleotides in length, as well as several sources of ssDNA, such as M13 and φX174 viral DNA and denatured calf thymus DNA, and obtained similar sensitivity. The OliGreen reagent has also been used to detect phosphodiester and phosphorothioate oligonucleotides.ref

Significant disadvantages of the UV absorbance method for oligonucleotide quantitation include the large relative contribution of free nucleotides to the signal and the interference caused by contaminants commonly found in nucleic acid preparations. By contrast, nucleotides and short oligonucleotides of six bases or less do not interfere with the OliGreen ssDNA quantitation assay. However, the OliGreen ssDNA quantitation reagent does exhibit fluorescence enhancement when bound to dsDNA and RNA. Like the PicoGreen assay, the linear detection range of the OliGreen assay in a standard fluorometer extends over four orders of magnitude — from 100 pg/mL to 1 µg/mL — with a single dye concentration (Figure 8.61). The linearity of the OliGreen assay is maintained in the presence of several compounds commonly found to contaminate nucleic acid preparations, including salts, urea, ethanol, chloroform, detergents, proteins, ATP and agarose (Table 8.14); however, many of these compounds do affect the signal intensity to some extent, so standard curves should be generated using solutions that closely mimic those of the samples. The OliGreen assay can even be performed using samples as complex as whole blood or serum.ref

Our experiments with homopolymers have demonstrated that the OliGreen reagent may exhibit significant base selectivity. The OliGreen reagent shows a large fluorescence enhancement when bound to poly(dT) but only a relatively small fluorescence enhancement when bound to poly(dG) and little signal with poly(dA) and poly(dC). Thus, it is important to use an oligonucleotide with similar base composition when generating the standard curve.

The remarkable properties of our OliGreen ssDNA quantitation reagent make it ideal for the fast and accurate detection and quantitation of:

  • Antisense oligonucleotides
  • Aptamers
  • Genomic DNA isolated under denaturing conditions
  • PCR primers
  • Phosphorothioate and phosphodiester oligodeoxynucleotides ref
  • Sequencing primers
  • Single-stranded phage DNA

Each vial of the OliGreen ssDNA quantitation reagent (O7582) contains a sufficient amount of dye for at least 200 assays using a 2 mL assay volume and a standard fluorometer, or 2000 assays using a 200 µL assay volume and a fluorescence microplate reader. The product is accompanied by a simple protocol that ensures linear and reproducible quantitation of ssDNA. We also provide the OliGreen reagent in a kit form — the OliGreen ssDNA Quantitation Kit (O11492) — that contains:

  • The OliGreen ssDNA quantitation reagent
  • A low-fluorescence, nucleic acid– and nuclease-free assay buffer concentrate, essential for the accurate measurement of ssDNA
  • An oligonucleotide standard (M13 sequencing primer) solution for assay calibration
  • A detailed protocol for ssDNA quantitation (Quant-iT OliGreen(R) ssDNA Reagent and Kit)

This kit provides sufficient reagents for 200 assays using a 2 mL assay volume and a standard fluorometer or 2000 assays using a 200 µL assay volume and a fluorescence microplate reader.

RiboGreen RNA Quantitation Assay

RiboGreen RNA Quantitation Reagent and Kit

The Patented RiboGreen RNA quantitation reagent (R11491) is our premier stain for quantitating RNA in solution. Like the PicoGreen dsDNA and OliGreen ssDNA quantitation assays, the RiboGreen RNA quantitation assay relies on a proprietary dye that exhibits a large fluorescence enhancement upon binding to nucleic acids. The extinction coefficient of the RiboGreen reagent, as well as its quantum yield and fluorescence enhancement upon binding RNA, are all significantly greater than those of ethidium bromide (Table 8.15).

The RiboGreen assay allows detection of as little as 1 ng/mL RNA in a standard fluorometer, fluorescence microplate reader or filter-based fluorometer using standard fluorescein excitation and emission settings (Figure 8.62). This sensitivity is at least 200-fold better than that achieved with ethidium bromide ref and at least 1000-fold better than that achieved using conventional absorbance measurements at 260 nm (an A260 of 0.1 corresponds to an ~4 µg/mL RNA solution). Unlike UV absorbance measurements at 260 nm, the RiboGreen reagent does not detect significant sample contamination by free nucleotides.ref Thus, the RiboGreen reagent more accurately measures the amount of intact RNA polymers in potentially degraded samples.

Using two different dye concentrations to cover its full dynamic range of three orders of magnitude, we have observed a linear correlation between the RNA concentration and fluorescence for 1.0 ng/mL to 50 ng/mL RNA using a 4000-fold dilution of the RiboGreen dye, and for 20 ng/mL to 1 µg/mL using a 400-fold dilution of the dye (Figure 8.62). Assay linearity is maintained in the presence of several compounds commonly found in nucleic acid preparations, including salts, urea, ethanol, chloroform, detergents, proteins and agarose ref (Table 8.16).

The RiboGreen RNA reagent is not appreciably selective for RNA — the dye also shows significant fluorescence enhancement upon binding to DNA. However, a simple DNase pretreatment of samples removes the contribution of DNA to the signal; see below for a description of our RiboGreen RNA-Specific Quantitation Kit with DNase I (R32702). The RiboGreen reagent may also have some base selectivity; it exhibits about 60% less fluorescence when bound to poly(G) homopolymers and virtually no fluorescence when bound to poly(U) or poly(C) homopolymers compared with the fluorescence when bound to poly(A) homopolymers or to rRNA.ref

Using the RiboGreen RNA quantitation reagent, we have reproducibly quantitated RNA from a wide variety of sources, including ribosomal RNA (rRNA), transfer RNA (tRNA), viral RNA, polyA+ fractions and total cellular RNA.ref The RiboGreen reagent is ideal for:

  • Fast and accurate measurements of RNA yields before generating cDNA
  • Determination of RNA yields from in vitro transcription
  • Accurate measurements of RNA before performing Northern blot analysis,ref S1 nuclease assays, RNase-protection assays, reverse-transcription PCR and differential-display PCR
  • Assay of DNA-dependent RNA polymerase activity ref
  • Detection of capillary electrophoresis–separated viral RNA that has been stained in vitro ref

Each vial of the RiboGreen RNA quantitation reagent (R11491) contains a sufficient amount of dye for at least 200 high-range assays or 2000 low-range assays using a 2 mL assay volume and a standard fluorometer. With a fluorescence microplate reader and a 96-well microplate, the assay volume is reduced to 200 µL, allowing 2000 high-range assays or 20,000 low-range assays. Included with each vial of the RiboGreen reagent is a simple protocol (Figure 8.57, Quant-iT RiboGreen(R) RNA Reagent and Kit) that ensures linear and reproducible quantitation of RNA. We also provide the RiboGreen reagent in a kit — the RiboGreen RNA Quantitation Kit (R11490) — that contains:

  • The RiboGreen RNA quantitation reagent
  • A low-fluorescence, nucleic acid– and nuclease-free assay buffer concentrate, essential for the accurate measurement of RNA
  • A ribosomal RNA standard (16S and 23S rRNA from Escherichia coli) solution for assay calibration
  • A detailed protocol (Quant-iT RiboGreen(R) RNA Reagent and Kit)

This kit provides sufficient reagents for at least 200 assays using a 2 mL assay volume and a standard fluorometer or at least 2000 assays using a 200 µL assay volume and a fluorescence microplate reader. The RNase-free TE buffer concentrate, which is essential to the success of the assay, is also available separately (T11493) and can be used to extend the number of low-concentration assays possible with the kit.

RediPlate 96 RiboGreen RNA Quantitation Kit

The RiboGreen assay is also available in a convenient RediPlate 96 RiboGreen RNA Quantitation Kit in which the RiboGreen reagent is predispensed into a 96-well microplate (R32700). The buffer and sample are simply added to the microplate wells — there is no need to handle the RiboGreen reagent. After a 10-minute incubation, the microplate is ready to read in a fluorescence microplate reader. The RediPlate RiboGreen assay has a linear range of ~15–1000 ng/mL (~3–200 ng in a 200 µL assay volume) with a single dye concentration (Figure 8.63). The microplate used in the RediPlate 96 RiboGreen RNA Quantitation Kit is provided in a resealable foil packet, and it snaps apart into twelve strips to permit assays in any multiple of eight (Figure 8.60). Eleven of the strips are preloaded with the RiboGreen reagent; the remaining strip, marked with blackened tabs, contains a series of RNA standards for generating a standard curve. In addition to the 96-well microplate, each RediPlate RiboGreen 96 RNA Quantitation Kit includes RNase-free reaction buffer and detailed instructions (RediPlate 96 RiboGreen(R) RNA Quantitation Kit). Multiple plates are available at significant discounts.

RiboGreen RNA-Specific Quantitation Kit with DNase I

Although the RiboGreen RNA quantitation reagent is insensitive to proteins, detergents, salts and other common contaminants, it is sensitive to both RNA and DNA. The presence of DNA in experimental samples will therefore elicit substantial errors in quantitative determinations of RNA. The RiboGreen RNA-Specific Quantitation Kit (R32702) is designed to circumvent this problem. The DNase I provided in this kit is used to reduce DNA concentrations from as high as 1 µg/mL to minimally detectable levels prior to fluorescence quantitation of RNA using the RiboGreen reagent. Because DNase I digestion and the RiboGreen fluorescence quantitation assay have quite different buffer-composition requirements, the two reactions must be carried out sequentially. The linear range for RNA quantitation with the RiboGreen RNA-Specific Quantitation Kit extends over almost three orders of magnitude in concentration — from 5 to 1000 ng/mL RNA (Figure 8.64). The lower limit of detection is about 40 times lower than that achieved with ethidium bromide–based fluorescence assays,ref and about 250 times lower than that achieved with A260 measurements. The original RiboGreen assay described above attains a slightly greater sensitivity due to the absence of DNase I and its attendant divalent cations. The RiboGreen RNA-Specific Quantitation Kit with DNase I is configured for use with a fluorescence microplate reader and contains:

  • The RiboGreen RNA quantitation reagent
  • A low-fluorescence, nucleic acid– and nuclease-free assay buffer concentrate, essential for the accurate measurement of RNA
  • DNase I from bovine pancreas
  • Concentrated DNase I buffer
  • A ribosomal RNA standard (16S and 23S rRNA from Escherichia coli) solution for assay calibration
  • A detailed protocol (Product Information Sheet)

Sufficient reagents are provided for at least 2000 assays using 200 µL assay volumes in a 96-well microplate format.

Other Stains for Nucleic Acid Quantitation in Solution

Cyanine Dyes and Phenanthridine Dyes for Nucleic Acid Quantitation in Solution

The dimeric cyanine dyes TOTO-1 and YOYO-1 are useful for sensitive fluorometric measurement of dsDNA, ssDNA and RNA in solution,ref although the PicoGreen, OliGreen and RiboGreen reagents generally are faster, have greater sensitivity and show a linear response over a broader range of nucleic acid concentrations. The linear range of assays that use the TOTO-1 and YOYO-1 dyes for DNA quantitation encompasses about two orders of magnitude, with a sensitivity limit of about 0.5 ng/mL. The TOTO-1, YOYO-1 and YO-PRO-1 nucleic acid stains have been used to quantitate PCR amplification products in a homogeneous human leukocyte antigen (HLA) typing method that requires no transfer or washing steps, thus minimizing the risk of sample contamination.ref Other dyes for nucleic acid quantitation in solution and their applications include:

  • YOYO-1 dye (Y3601, Section 8.1) for solution quantitation of oligonucleotides,ref PCR products ref and nuclear run-on assays ref
  • YO-PRO-1 dye (Y3603, Section 8.1) for quantitating dsDNA in a fluorescence microplate reader, with a reported sensitivity limit of about 2.5 ng/mL ref
  • YO-PRO-1 dye for quantitating RNA isolated from Xenopus embryos ref
  • YO-PRO-1 dye and SYBR Green I dye for direct counting of viruses in marine and freshwater environments ref (photo)
  • PO-PRO-3 dye (P3585, Section 8.1) for quantitating DNA in a fluorescence microplate reader,ref with fluorescence measurements reported to be independent of base-pair composition
  • Ethidium bromide (E1305, E3565; Section 8.1) for measuring the yield of PCR products ref
  • Ethidium bromide and acridine homodimer (A666, Section 8.1) for quantitating covalently closed, circular DNA and for measuring the activity of polymerases, deoxynucleotidyl transferases, ligases, gyrases, topoisomerases and nucleases ref
  • YOYO-1 dye for measuring the activity of DNases ref
  • SYBR Gold dye (S11494, Section 8.4), for detecting DNA mutations in a simple PCR-based assay ref
  • SYBR Green I dye for quantitation of nucleic acids on plastic wrap or paraffin (Product Highlight: SYBR Green I and SYBR Green II Dyes for Quantitating Nucleic Acids on Plastic Wrap or Paraffin Sheets)

Dyes such as the ethidium homodimers and our dimeric cyanine dyes — the TOTO, YOYO, BOBO, POPO, JOJO and LOLO dyes (Table 8.2) — exhibit a high affinity for double-stranded nucleic acids but label small single-stranded oligonucleotides less well. This characteristic of ethidium homodimer-1 (E1169, Section 8.1) was exploited to analyze short self-annealing oligonucleotides for their ability to hybridize.ref Because our dimeric cyanine dyes and the PicoGreen dsDNA quantitation reagent have extremely low intrinsic fluorescence in the absence of DNA, high fluorescence enhancements upon binding, higher quantum yields and much larger extinction coefficients than ethidium homodimer-1,ref they should prove superior in this application.

Hoechst 33258 Dye for Quantitating DNA in Solution

The Hoechst 3325